The Model 106 pulse generator allows for tracking of the patient’s heart rate patterns, and includes a programmable feature that triggers stimulation when the heart rate exceeds a physician-selected threshold. To address variability between patients, this programmable feature allows the detection threshold to be tailored to the individual patient. The optional feature (“AutoStimulation” or "Autostim Mode”) will serve its intended utility in situations where a seizure is associated with a heart rate increase. Finally, the design of the feature includes a risk mitigation to prevent excessive duty cycle (ON time > OFF time). At the time of implant, the physician will confirm that the implanted components accurately detect the heart rate, adjusting the heartbeat sensing settings if necessary. Initial detection algorithm settings will be selected for the algorithm performance adjustment phase. At the time of physician follow-up, the physician will decide if the algorithm should be adjusted to provide more or fewer automated responses based on the patient reports of seizure frequency, the device log, and to balance the battery drain consequence of providing unneeded extra VNS automatic magnet mode output.

Discussion

Enote Oct 1, 2019:
Лишний рабочий цикл В сабже речь не о циклах, а о соотношения времен Вкл/Откл внутри цикла (периода). Перевести сабж можно по-разному, например, "чрезмерная ПВ", но это точно не лишний рабочий цикл.

Proposed translations

2 hrs
Selected

для предотвращения излишнего рабочего цикла

*
Something went wrong...
4 KudoZ points awarded for this answer.
3 hrs

предотвращение чрезмерной скважности импульсов

Не знаю как в медицине, но в технике для описания соотношения между длительностями состояний Вкл и Откл периодических импульсов используется понятие Скважность и близкие понятия Коэффициент заполнения и Продолжительность включения (ПВ). Имхо, скважность вполне подходит для контекста
Something went wrong...

Reference comments

2 days 9 mins
Reference:

чрезмерная напряженность рабочего цикла (<50%)

http://www.cardion.cz/file/473/VNS Therapy System Physician&...
The percentage of time the pulse generator is stimulating is called a
“duty cycle.” A duty cycle is calculated by dividing the stimulation
time (programmed ON time plus 2 seconds of ramp-up time and
2 seconds of ramp-down time) by the sum of the ON and OFF
times.

https://us.livanova.cyberonics.com/healthcare-professionals/...
DUTY CYCLE: 10%
ON Time 30 sec
OFF Time 5 min
Frequency 30 Hz
Pulse Width 500 μsec

--------------------------------------------------
Note added at 2 days 10 mins (2019-09-28 11:48:06 GMT)
--------------------------------------------------

поправка: >50%